Circuit Protection & ControlCircuit Protection & ControlFlow Meter FC01-LQUser handbookFlow Meter FC01-LQUser handbookM_FC01_LQ_0408_eM_FC01_L

Page 2

Flow Meter FC01-LQ Flow Meter FC01-LQ4 Description 4 DescriptionThe power supply is physically isolated between power supply input and system powersup

Page 3 - Table of content

Flow Meter FC01-LQ Flow Meter FC01-LQDescription 5 Description 5Power supply: DC 24 V supply Internal switch mode power supply with physical isolation

Page 4

Flow Meter FC01-LQ Flow Meter FC01-LQ6 Installation 6 Installation2 InstallationThese are general directions for the application of calorimetric measu

Page 5

ø20ø1818.214648ø24retention slotO ringmonitoring head CSP-11..O ringretention pinunion nutsensor adapter TP-…Flow Meter FC01-LQ Flow Meter FC01-LQInst

Page 6

Flow Meter FC01-LQ Flow Meter FC01-LQ8 Installation 8 Installation2.2.1.1 Sensoradapter TP-..The sensor adapter TP-.. is available in 6 pipe diameters

Page 7 - 1 Description

41505470SW8810011013121242428tød25324050DNTypeBV-03M3BV-04M3BV-05M3BV-06M325324050G1"11/4"11/2"2"L59657785H115115150150AHødGtSWALF

Page 8

monitoring head should be alignedin direction of flow (see arrow)14L14ø18ø22SW20Typ eLCSF-…L43… 190CSF-…L30… 300CSF-…L40… 400M16x0.75round connectorfi

Page 9 - 1.2 System description

Flow Meter FC01-LQ Flow Meter FC01-LQInstallation 11 Installation 112.2.2.1 Mounting instructions for monitoring head CSF-11Caution!The two sensors (M

Page 10 - 1.2.1 User interfaces

Flow Meter FC01-LQ Flow Meter FC01-LQ12 Installation 12 Installation● The monitoring head should be installed only in a straight section of piping. T

Page 11

Locking set 011 chain 4 x 32 DIN 5685 (approx. 1 m)2 catch for chain NG 53 clip with screws and nuts DN15 to DIN 11850(tightening torque 10 Nm)321Flow

Page 12 - 2 Installation

Please follow these installation and adjustment instructions carefully. Failure to comply withthese instructions or misuse of this equipment will void

Page 13

Flow Meter FC01-LQ Flow Meter FC01-LQ14 Installation 14 Installation2.3 Installation of electronic control unit FC01-LQ2.3.1 Mechanical installation2.

Page 14

Flow Meter FC01-LQ Flow Meter FC01-LQInstallation 15 Installation 152.3.1.2 Surface mounted version FC01-LQ-FH-U1... ● Remove the cover of the housing

Page 15

Flow Meter FC01-LQ Flow Meter FC01-LQ16 Installation 16 Installation2.3.1.3 Front panel mounted housing FC01-LQ-ST-U1... ● Insert housing in the front

Page 16

Flow Meter FC01-LQ Flow Meter FC01-LQInstallation 17 Installation 172.3.2 Electrical connectionValid for all plug-in screw terminal strips:Cable size

Page 17

Flow Meter FC01-LQ Flow Meter FC01-LQ18 Installation 18 InstallationXAO - - Analogue outputsConnection by 8 pole connector; max. 1.5 mm2; LiYCY 2 x 0.

Page 18

Flow Meter FC01-LQ Flow Meter FC01-LQInstallation 19 Installation 195 LIM2 E limit value 2 - emitter terminal6 LIM2 C limit value 2 - collector termin

Page 19

Flow Meter FC01-LQ Flow Meter FC01-LQ20 Installation 20 Installation2.3.2.1 Circuit diagram FC01-LQVersion: 24 V, relay outputs2.3.2.1 Circuit diagra

Page 20 - 2.3.1 Mechanical installation

Flow Meter FC01-LQ Flow Meter FC01-LQInstallation 21 Installation 212.3.2.2 Circuit diagram FC01-LQVersion: 24 V, transistor outputs (NPN)2.3.2.2 Cir

Page 21 - FC01-LQ-FH

Flow Meter FC01-LQ Flow Meter FC01-LQ22 Installation 22 Installation2.3.2.3 Electrical connection - frequency output (version FC01-LQ-U1T4)The quantit

Page 22

Flow Meter FC01-LQ Flow Meter FC01-LQInstallation 23 Installation 23Electromechanical pulse counter (fig. 18)The FC01-LQ driver output comprises an in

Page 23 - 2.3.2 Electrical connection

Flow Meter FC01-LQ Flow Meter FC01-LQTable of contents I Table of contents ITable of content1 Description . . . . . . . . . . . . . . . . . . . . . .

Page 24

3 Operating systemClear menu-driven control, via keyboard and display, enables easy definition of parameters andconfiguration. This provides high syst

Page 25

Flow Meter FC01-LQ Flow Meter FC01-LQOperating system 25 Operating system 25Menu pagingThe next menu option is selected by pressing MODE (forward pag

Page 26 - Flow Controller

4 Operation and main menu4.1 Switch-on performanceUpon power application POWER ON TEST will be shown on the display for approx. 1 sec., withthe softwa

Page 27 - Flow Monitor

Flow Meter FC01-LQ Flow Meter FC01-LQOperation and main menu 27 Operation and main menu 274.2.1 Operating data4.2.1.1 Measured value(s)Flow rate and m

Page 28

Flow Meter FC01-LQ Flow Meter FC01-LQ28 Operation and main menu 28 Operation and main menuThe limit switches are identified according to their physica

Page 29

Flow Meter FC01-LQ Flow Meter FC01-LQOperation and main menu 29 Operation and main menu 294.2.1.4 Main menu 4.2.1.4 Main menudeletemaximum values▲ + ▼

Page 30 - 3 Operating system

5 Configuration (menu option: CONFIGURATION)The CONFIGURATION menu serves to adjust the FC01-LQ to the application within the entiresystem. During sys

Page 31

Flow Meter FC01-LQ Flow Meter FC01-LQConfiguration 31 Configuration 31Enter the following characteristics when selecting a CSP monitoring head with se

Page 32 - 4 Operation and main menu

Flow Meter FC01-LQ Flow Meter FC01-LQ32 Configuration 32 Configuration5.4 Flow rate unit (menu option: FLOW UNIT)At these point (1st line top left) th

Page 33 - 4.2.1 Operating data

Caution!The content of the totalizer is deleted in the event of power failure or disconnection of thepower supply!Skipping to the submenu BARGRAPH or

Page 34

Flow Meter FC01-LQ Flow Meter FC01-LQII Table of contents II Table of contents4 Operation and main menu . . . . . . . . . . . . . . . . . . . . . .

Page 35

Flow Meter FC01-LQ Flow Meter FC01-LQ34 Configuration 34 Configuration5.8 Pipe diameter (menu option: PIPE SIZE)If flow volume/time unit display has b

Page 36

Flow Meter FC01-LQ Flow Meter FC01-LQConfiguration 35 Configuration 355.10 Analogue output - flow rate (menu option: ANA OUT FLOW)This menu option all

Page 37

Flow Meter FC01-LQ Flow Meter FC01-LQ36 Configuration 36 Configuration• ERR. BARGRAPH OUT OF RANGE(bar value outside measuring range) • ERR. BARGRAPH

Page 38

Flow Meter FC01-LQ Flow Meter FC01-LQConfiguration 37 Configuration 375.13 Configuration menu5.13 Configuration menuConfig. plausible?yessubmenutemper

Page 39

Flow Meter FC01-LQ Flow Meter FC01-LQ38 Configuration 38 Configurationor▲▼submenusensor selectBV SIZE SELECTDN25 (1 in)BV SIZE SELECTDN32 (1.1/4 in)BV

Page 40

Flow Meter FC01-LQ Flow Meter FC01-LQConfiguration 39 Configuration 39submenuflow unitsubmenutemperature unitFLOW UNITMETRE/SEC (m/s)TEMP. UNITCELSIUS

Page 41

Flow Meter FC01-LQ Flow Meter FC01-LQ40 Configuration 40 Configurationsubmenudisplay selectionTotalizer function?or▲▼FREQUENCY OUTPUT?M → yes other →

Page 42

Flow Meter FC01-LQ Flow Meter FC01-LQParameter selection 41 Parameter selection 416 Parameter selection (menu option: PARAMETERS)After configuration o

Page 43 - 5.13 Configuration menu

Flow Meter FC01-LQ Flow Meter FC01-LQ42 Parameter selection 42 Parameter selectionExample for ON:FC01-LQ with relay outputs (option R2):• LIM1 - LIM1

Page 44 - 5.14 Configuration submenus

Flow Meter FC01-LQ Flow Meter FC01-LQParameter selection 43 Parameter selection 436.5 Quitting the parameter selection menuBefore the parameter select

Page 45

Flow Meter FC01-LQ Flow Meter FC01-LQTable of contents III Table of contents III8 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 46 - ° US-GALLONS

Flow Meter FC01-LQ Flow Meter FC01-LQ44 Parameter selection 44 Parameter selection6.6 Parameter selection menu 6.6 Parameter selection menuEND OF PA

Page 47

Flow Meter FC01-LQ Flow Meter FC01-LQErrors 45 Errors 457 Errors7.1 Test and diagnosisThe FC01-LQ is provided with extensive test and diagnosis functi

Page 48

Flow Meter FC01-LQ Flow Meter FC01-LQ46 Errors 46 Errors7.2 Potential errorsIndependent of the priority group, all errors found are indicated with the

Page 49

Flow Meter FC01-LQ Flow Meter FC01-LQErrors 47 Errors 47Priority group IIIError Cause RectificationNo. 20 Medium temperature too lowNo. 30 Over limits

Page 50 - Paramet. plausible?

Flow Meter FC01-LQ Flow Meter FC01-LQ48 Technical data 48 Technical data8 Technical data8.1 Ambient conditionsrail-mounted surface mounted front panel

Page 51 - 7 Errors

Flow Meter FC01-LQ Flow Meter FC01-LQTechnical data 49 Technical data 49Inrush current: Ip= typ. 3 A (20 µs)Switch-off current: Ikipp= typ. 0.75 ARate

Page 52 - 7.2 Potential errors

Flow Meter FC01-LQ Flow Meter FC01-LQ50 Technical data 50 Technical data8.3.1 Voltage output V1 - 5 V FSSignal voltage range: US= 0 V (1 V) to 5 V ±2%

Page 53

Resistive loadMax. admissible switching capacity: 50 WMax. admissible switching current: 1 AMax. admissible continuous current: 1 AMax. admissible swi

Page 54 - 8 Technical data

Flow Meter FC01-LQ Flow Meter FC01-LQ52 Technical data 52 Technical dataResistive loadMax. admissible switching capacity: 1.5 WMax. admissible switchi

Page 55 - 8.3 Analogue outputs

Flow Meter FC01-LQ Flow Meter FC01-LQTechnical data 53 Technical data 538.5 Metrological data 8.5.1 Flow rate measurementMeasuring is possible up to t

Page 56 - 8.4 Signal outputs

Flow Meter FC01-LQ Flow Meter FC01-LQ

Page 57 - 8.4.2 Transistor outputs (DC)

Flow Meter FC01-LQ Flow Meter FC01-LQ54 Technical data 54 Technical dataInner pipe diamater Measuring range Display rangeD in mm in m3/h in m3/h50 21

Page 58

8.5.2 Temperature measurement:Measuring range: -40 °C … +130 °CAccuracy: ±1% MB ***8.5.3 Electronic control unit FC01-LQ Temperature drift: 0.1%/K/MBE

Page 59 - 8.5 Metrological data

Flow Meter FC01-LQ Flow Meter FC01-LQ56 Technical data 56 Technical data8.6 Sensor interface 8.6.1 Electrical data of the terminal for calorimetric mo

Page 60

9 AccessoriesNo. Accessory Ordering configuration1 surface mounted housing FC01-LQ-FH2 front mounted housing FC01-LQ-ST3 connecting cable for calorim

Page 61

Flow Meter FC01-LQ Flow Meter FC01-LQ

Page 62 - 8.6 Sensor interface

Flow Meter FC01-LQ Flow Meter FC01-LQIndex IndexIndexAANA OUT FLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 63 - 9 Accessories

Flow Meter FC01-LQ Flow Meter FC01-LQIndex IndexMMain menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 64

Flow Meter FC01-LQ Flow Meter FC01-LQAppendix 1 Appendix 1MAXMINMINMINMINMINMINMINXMINMINMINXMINXXXXONOFFOFFOFFOFFOFFOFFOFFONOFFOFFOFFONOFFONFAONONONO

Page 65

Flow Meter FC01-LQ Flow Meter FC01-LQ

Page 66

Flow Meter FC01-LQAppendix 2Appendix 2 - Menu structure of the FC01-LQ (operator dialog)not available with sensor type CSPnur bei gewählter Totalisato

Page 67

Flow Meter FC01-LQ Flow Meter FC01-LQDescription 1 Description 11 DescriptionThe Flow Meter FC01-LQ is designed to detect flow speed, flow volume and

Page 68

Flow Meter FC01-LQ Flow Meter FC01-LQ2 Description 2 Description1.1 Calorimetric monitoring procedureThe calorimetric measuring procedure is based on

Page 69

Flow Meter FC01-LQ Flow Meter FC01-LQDescription 3 Description 31.2 System descriptionThe system comprises the following hardware functional modules:1